This Sensor Cartridge with Infrared Sensor Options comes with - 4 gas -
Oxygen, %VOL CH4 IR, Carbon Monoxide (CO), Sulfur Dioxide (SO2) for Impact Pro
IR Multi-Gas Detector
Infrared
sensors for flammable gases cannot detect hydrogen (H2). If the
environment is suspected of containing combustible levels of hydrogen,
a cartridge that uses a standard catalytic bead sensor should be used
with the Impact Pro instead.
The flammable (%LEL or %V/V) IR
sensors are factory calibrated to respond linearly to methane (CH4).
Response to other hydrocarbons is non-linear and the Impact Pro IR will
not display corrected values for other hydrocarbon gases. The Impact
Pro IR will generally (but not always) give a higher reading for heavy
hydrocarbons than for methane.
All IR sensor cartridges require
a modified Impact Pro IR cartridge bay, which is designated as
modification state 15. If not present, this modification must be
performed by your authorized Honeywell Analytics Service Center.
Failure to do this may lead to physical damage when attempting to
insert the new IR sensor cartridge and will not be covered by warranty.
Any electrochemical sensors in this cartridge cannot operate accurately
in oxygen deficient environments and should not be used for continuous
monitoring in inert atmospheres.